dc.description.abstractTansiyonun ölçülmesi ve sürekli izlenmesi, hipertansiyon başta olmak üzere bir çok kalp hastalığının erken teşhis edilmesinde çok önemli rol oynar. Bu çalışmada kelepçesiz tansiyon aleti analizi ve tasarımı amaçlanmıştır. Kelepçesiz tansiyon ölçümünün temel çalışma prensibi atımlı geçiş süresine (PTT) dayanır. Atımlı geçiş süresi (PTT), elektrokardiyografinin (EKG) R-tepe noktası ile fotopletismogram (PPG) sinyalinin maksimum değer yüksekliğinin %25'i veya %50'si olan nokta arasındaki süredir. PTT değeri ile tansiyon arasındaki ilişki kullanılarak regresyon analizi ile tansiyon değerleri tahmini yapılır. Literatür de yapılan çalışmalar detaylı bir şekilde incelenmiştir ve eksik görülen noktalar üzerinde çalışılmıştır. Literatürdeki çalışmalardan farklı olarak atımlı geçiş zamanı (PTT) için birden fazla değer ele alınarak çalışmalar yürütülmüştür. Bu çalışmada EKG, PPG ve tansiyon değerlerinden oluşan hazır veriler kullanılarak 4000 hastanın verileri incelenmiştir. PTT değeri hesaplanırken literatürde en çok kullanılan dört nokta belirlenmiştir ve bu dört noktadan PTT hesaplanması yapılmıştır. PTT hesaplandıktan sonra farklı regresyon analiz yöntemleri kullanılarak tansiyon sonuçları elde edildi ve en başarılı yöntemi belirlenmesi için analiz çalışması yapılmıştır. Elde edilen tansiyon tahmin sonuçları daha sonra Tıbbi Cihaz Geliştirme Derneği (AAMI) ve İngiliz Hipertansiyon Derneği (BHS) standartları dikkate alınarak uygunluklarına göre sınıflandırılmıştır. Çalışmanın ikinci aşamasında cihaz prototipi oluşturulmuştur. EKG kiti ve PPG sensörü ile veriler alınarak analiz edilmiştir. Sensörler ile veriler alınırken, eş zamanlı olarak kalibrasyonu yapılmış bir tansiyon aleti ile tansiyon değerleri ölçülmüştür. Bu değerler referans tansiyon değerleri olarak kullanılmıştır. Sensörlerden alınan verilerden PTT değerleri hesaplanarak regresyon analizi yöntemleri ile tansiyon tahmin sonuçları elde edilmiştir. Sensörlerden alınan verilerden elde edilen sonuçlar ile hazır verilerden elde edilen sonuçlar karşılaştırılmıştır. Prototip cihaz, sistolik kan basıncı (SBP), diyastolik kan basıncı (DBP) ve ortalama arter basıncı (MAP) için en doğru sonucu veren yöntemler tek bir PTT değeri ve regresyon yöntemi kullanmak yerine farklı yaklaşımlar ele alınmıştır. Hazır verilerle yapılan analizin sonucunda sistolik kan basıncı (SBP), diyastolik kan basıncı (DBP) ve ortalama arter basıncı (MAP) tahmin sonuçları elde edilirken farklı regresyon yöntemleri ve PTT değerleri kullanılmasına karar verilmiştir. Farklı yöntemler kullanılarak doğruluğu daha yüksek bir cihaz prototipi oluşturulmuştur.en_US
dc.description.abstractMeasurement and continuous monitoring of blood pressure plays a significant role in the early diagnosis of many heart diseases, especially hypertension. The aim of this study was to analyze and develop cuffless blood pressure monitors. The basic principle of operation of blood pressure measurement without cuff is based on pulse transition time (PTT). Pulse transition time (PTT) is the time between the R peak of the electrocardiogram (ECG) and the point at which the photopletismogram (PPG) signal is 25% or 50% of the maximum value. Blood pressure values are estimated by regression analysis based on the relationship between the PTT value and blood pressure. The studies performed in the literature were reviewed in detail and the missing points were investigated. Unlike the studies in the literature, studies were conducted in which more than one value for pulse transition time (PTT) was considered. In this study, data from 4000 patients were examined, using preformed data consisting of ECG, PPG, and blood pressure values. In calculating the PTT value, the four most commonly used points in the literature were determined, and PTT was calculated from these four points. After calculating PTT, blood pressure values were obtained and analyzed by different methods of regression analysis to determine the most significant method. The obtained blood pressure prediction results were then classified according to their compliance with the Medical Device Development Association (AAMI) and British Hypertension Association (BHS) standards. In the second phase of the study, a prototype of the device was developed. Data were analyzed using an ECG kit and a PPG sensor. While data were recorded with the sensors, blood pressure values were measured with a blood pressure monitor, which was calibrated at the same time. These values were used as reference blood pressure values. PTT values were calculated from the data recorded by the sensors, and blood pressure prediction results were obtained by regression analysis techniques. The results from the data recorded by the sensors and the results from the precalculated data were compared. The prototype of the device, the methods that provide the most accurate results for systolic blood pressure (SBP), diastolic blood pressure (DBP), and mean arterial pressure (MAP), instead of using a single PTT value and a regression method, different approaches are discussed. As a result of the analysis with prefabricated data, it was decided to use different regression methods and PTT values, obtaining the prediction results for systolic blood pressure (SBP), diastolic blood pressure (DBP), and mean arterial pressure (MAP). Different methods were used to develop the prototype of a device with higher accuracy.en_US
dc.language.isoturen_US
dc.publisherFatih Sultan Mehmet Vakıf Üniversitesi, Lisansüstü Eğitim Enstitüsüen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectElektrokardiyografi (EKG)en_US
dc.subjectFotopletismogram (PPG)en_US
dc.subjectAtımlı Geçiş Süresi (PTT)en_US
dc.subjectTansiyonen_US
dc.subjectRegresyon Analizien_US
dc.subjectSensören_US
dc.subjectElectrocardiography (ECG)en_US
dc.subjectPhotopletismogram (PPG)en_US
dc.subjectPulse Transition Time (PTT)en_US
dc.subjectBlood Pressureen_US
dc.subjectRegression Analysisen_US
dc.subjectSensoren_US
dc.titleKelepçesiz Tansiyon Aleti Analizi ve Tasarımıen_US
dc.title.alternativeAnalysis and Design of Cuff-Less Blood Pressure Deviceen_US
dc.typemasterThesisen_US
dc.contributor.departmentFSM Vakıf Üniversitesi, Lisansüstü Eğitim Enstitüsü, Biyomedikal Mühendisliği Ana Bilim Dalıen_US
dc.relation.publicationcategoryTezen_US
dc.contributor.institutionauthorKaraçay, Sena Nur Acar


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record